Infrared thermometer with -50°C to 500°C measurement range

Infrared vs Probe Thermometers: Measuring the Right Temperature in the Lab

← Viking Lab Supply Learning Hub

Compare surface and liquid measurements, understand thermowells, and avoid confusing temperature range with accuracy.

Define what you need to measure

Before choosing a thermometer, identify the temperature that matters: an equipment surface, a liquid at a particular depth, or another defined location. Two instruments can display different temperatures while both are responding correctly to different parts of a system.

A heater setting is not automatically the sample temperature. Likewise, the outside of a beaker can differ from the liquid inside. Define the measurement location, required accuracy, expected range and response time before selecting an instrument.

Compare infrared and contact instruments

Instrument Typical measurement Main selection questions
Infrared thermometer Surface temperature without contact Emissivity, target size, distance and calibration
Contact probe Temperature at the sensing region Immersion requirements, compatibility, accuracy and response
Probe with thermowell Temperature through a protective interface Probe fit, insertion, added response lag and compatibility

The infrared thermometer listed from −50°C to 500°C is intended for surface checks. Its broad range does not establish a particular numerical accuracy. The laser is an aiming aid, not the sensing element. Follow its instructions for measuring distance and laser precautions.

The digital probe thermometer listed from −50°C to 300°C offers a contact approach where its probe and documented limits suit the application. The current listing does not document an immersion-depth or ingress-protection rating for the whole instrument, so do not infer that its display housing can be submerged.

Handheld infrared thermometer for non-contact surface measurements
Handheld infrared thermometer for non-contact surface measurements. View product details.

Why the readings may differ

Three locations, potentially three temperatures
1
Heater control
Selected setting or sensor reading
2
Vessel wall
Local outside surface temperature
3
Liquid sample
Temperature at the probe location

Conceptual measurement illustration. These values need not be equal; no numerical differences are implied.

Infrared readings depend on surface properties and the instrument's emissivity assumptions. Shiny surfaces can produce misleading results. The sensing spot also grows with distance, so a small target may not fill the measurement area. A laser dot should not be treated as the complete sensing footprint.

Do not aim through ordinary glass and interpret the reading as the temperature of the liquid behind it. For a liquid measurement, use a suitable contact arrangement and follow the probe instructions for insertion depth, clearance and stabilization. Avoid touching the vessel bottom if the procedure calls for measuring the liquid itself.

What a thermowell does

A thermowell provides a protective interface for a suitable sensor. It is not a thermometer and does not supply a calibration or accuracy rating. Select the well for the apparatus geometry and sensor fit; extra material between sample and sensor can change response time.

Compare the 45 mm stem thermowell, 100 mm stem thermowell and 200 mm stem thermowell. Their listed 24/40 joints do not eliminate the need to check internal clearance, probe dimensions and separation from moving stir bars.

Borosilicate thermowell with a 100 mm stem and 24/40 joint
Borosilicate thermowell with a 100 mm stem and 24/40 joint. View product details.

A practical selection checklist

Record the target location, range, required uncertainty, calibration status, material compatibility and response needs. Compare the exact instrument manual with those requirements. Neither a product title nor a display resolution proves the measurement accuracy you need.

Can I use an industrial infrared thermometer for body temperature?

Use an instrument specifically intended and validated for that medical purpose. The VLS industrial infrared listing is not intended as a medical thermometer.

What else helps a temperature setup?

Appropriate boss heads and laboratory clamps can support compatible apparatus. Read the support guide and joint compatibility guide before selecting connections.

Technical references

Fluke: infrared and contact measurement principles; Fluke: infrared measurement and glass. These references explain principles; they do not establish specifications for VLS instruments.